All data overview

Expert Vibro

Inputs / Outputs

Analog inputs4, 8, 12 or 16
Sampling rate, adjustable per channel20 Hz .. 50 kHz
Usable signal rangeDC .. 20 kHz
Voltage range± 25 V 
Current ranges0 .. 20 mA / 4 .. 20 mA / free
Signal conditioning, software switchableNone, AC coupling, IEPE
Resolution / input impedance24 bit / 4 MΩ
Dielectric withstand voltage± 100 VDC
Galvanic isolation: channel to channel / channel to system± 400 VDC
Measurement accuracy0.5 mV + 0.008 % from measurement value
Digital inputs4
Input signallow: 0 .. 1 V / high: 5 .. 100 VDC@3.5 mA 
Galvanic isolation: channel to other channel types / channel to system / PE± 400 VDC
Max. input frequency / min. pulse width1 MHz / 500 ns
Analog outputs4
Resolution16 bit
Output ranges0 .. 10 V / ± 10 V / 0 .. 20 mA / 4 .. 20 mA / ± 20 mA
Galvanic isolation: channel to other channel types / channel to system / PE± 400 VDC
Min. load / max. burden650 Ω
Digital outputs8
Max. switching voltage / current50 V / 2.5 A for 2 seconds, 250 mA continuous current
Galvanic isolation: channel to other channel types / channel to system / PE± 400 VDC
PWM basic frequency / duty cycle5 Hz .. 10 kHz / 1:1000

Data storage

Internal / external
2 .. 14 GB (approximately 30 million measurements per GB) / USB, NFS, CIFS, (S)FTP

Signal comditioning functions

High-pass / low-pass / bandpass filter
Cut-off frequency / filter order / filter characteristic5 .. 20 kHz / 4, 6, 8, 10 / incl. Butterworth
Integrator / Differentiation
Single or double integrator / differentiator
FFT
Line number / window function / averaging 400 .. 12.800 lines / von-Hann, Hamming, Flat-Top / 2 .. 32 times
FFT typesNarrow band / wide band, envelope / demodulation, amplitude / phase spectra
Characteristic values from time signal
Maximum / minimum value, peak-to-peak value, arithm. average, TRMS, max. vect. sum, local min. / max. values
Characteristic values from frequency spectrum
Amplitude, frequency and phase of main oscillation and any harmonics, sum value, quadratic mean (in any frequency band), value remainder
Characteristic values on digital input
Frequency 0.2 Hz .. 1 MHz
Counter 64 Bit, UP-Down counter, quadratur decoder

Interfaces

Physical equipment COM 1 / COM 2RS485, 9-pole sub-D-socket, DIN EN ISO 19245-1
Physical equipment COM 3RS232, 9-pole sub-D-plugs
LAN1 x 1000Base-TX
WLAN / WWAN (optional)802.11b/g/n / GPRS, UMTS, LTE max. 100 Mbit/s
USBDevice 2.0 / host 2.0 / low / high / full
PROFIBUS2 x PROFIBUS DPV1 / Slave max. 12 Mbit, also redundant according to PNO 2.212 V1.2
CAN / RS 232/4852 x CAN 2.0 / Modbus RTU, SCPI, ASCII
TCP/IPModbus TCP, OPC UA

General technical information

Dimensions / weightW 217 mm x H 88 mm x D 128 mm / 750 g
FixingRail mounting DIN EN 60715 or screw fixing
Signal connections Plug-in screw terminals, 96 terminals in 2 rows, max. 1,5 mm²
Temperature range-20 .. 50 °C*
Power supply 12 .. 24 VDC / ± 10%
Power inputmax. 20 W

* Sufficient air circulation in the control cabinet must be ensured.

Registered: 11.04.2019