All data overview

TopMessage / TopLab

Analog inputs
Voltage range± 156 mV ... ± 10 V
Current range0/4 ... 20 mA
Thermocouplesany, all types, integrated temperature compensation; resistance thermometer Pt100(0), NTC and linear resistance to 10 kΩ (not ADVT)
Potential isolation750 VDC for system and supply; 400 VDC between channels at ADGT module, 110 VDC between channels in other modules
Resolution24 Bit (≈7 decimal places) precision: V, mA 0,01 % from accumulated value Pt100 0.1 K; Pt1000 0.1 K thermocouple 0.1% from accumulated value
Analog outputs
Resolution16 Bit
Potential isolation750 V
Output signal0/4 ... 20 mA at maximum burden 650 Ω
Digital inputs
Potential isolation2.5 kV
Output signalslow: 0 ... 1.5 VDC@0 ... 1.5 mA / high: 3.5 ... 90 VDC@2 mA
Frequency / counter inputs
Potential isolation2.5 kV
Measurement rangelow: 0 ... 1.5 VDC@0 ... 1.5 mA / high: 3.5 ... 90 VDC@2 mA
Measurement frequencyup to 30 kHz to TTL-level
Digital outputs
Potential isolation2.5 kV
Switching voltagemax. 50 VDC@2.5 A
Data storage
Standard size /
measurement data
Storage partitionable, data compression standard: 500 kB; max. 96.000 measurement values
Maximum /
measurement data
1 GB; up to 128 millions measurement values
Interfaces
SerialRS232/485/422, switchable, PROFIBUS DP Slave optional
Physical equipment COM 19-pole sub-D connector, DIN EN ISO 19245-1
Protocolscustomer specific, modem connector, Modbus RTU Master / Slave
Physical equipment COM 29-pole sub-D connector, DIN 41652, chapter 1 (ISO 4902)
EthernetRJ45 (8-pole STP-socket), 10 BaseT (Twisted Pair, 10 Mbps, half-duplex), Protocols: TCP/IP, HTTP, SMTP, SNTP, Modbus TCP Client / Server
Modulbus
Physical equipment2-pole Phoenix plugs; internal bus for connecting additional modules
Baud rate1 MBaud (adjustable)
Lengthup to 10 m (1 MBaud)
General technical information
TopMessage dimensions200 x 73 x 118 mm
TopMessage weigth1 kg
TopMessage mountingRail mounting DIN EN 50023 or screw fixing
TopMessage signal connectionsDeatchable screw terminals, 33 terminals (2-rows), lead protection, connector cable, max. 2,5 mm²
TopLab dimensions226 x 145 x 180 mm
TopLab weight1 kg
TopLab signal connectionsup to 64 4 mm lab plugs, gold plated
Temperature range-20 .. 50 °C
Power supply12-36 VDC / 12-28 VAC eff. / ± 10%, for AMDT 18 VAC/DC, power input for master device: < 10 Watt
LC-DisplayFor servicing purposes: 4 lines à 15 characters 4 cursor keys, Enter, Esc, Reset keys, Displays configuration data, channel name, scaled and lineared measurement data

Registered: 25.05.2016